Rotronic CP11 CO2 Gas Measurement Instrument Meter

Rotronic CP11 CO2 For Measurement & Logging Of CO2 (Carbon Monoxide) Gas

Rotronic CP11 CO2 Measurement Data Logger & Instrument – Carbon Dioxide (CO2) Gas Measurement

Rotronic CP11 CO2 is a carbon dioxide measurement instrument – the economical multiple parameter meter for measuring and recording levels of COgas also calculates dew point and wet bulb temperature.

The Rotronic CP11 instrument is equipped with the field-tested Rotronic Hygromer IN-1 humidity sensor offering unbeatable value for money.

Typical applications of the Rotronic CP11 include air quality monitoring in underground car parks for safety and leak monitoring in filling plants.

Rotronic CP11 CO2 GAS Meter

Product Benefits

Key Features – The Rotronic CP11 is capable of calculating dew point and wet bulb temperature and features automatic or manual logging. There is also easy field calibration by simple connection to a known reference.

Connections – The CP11 features a mini USB port for connection to a PC  and a 5 VDC power supply, for connection of the optionally available AC adapter.

Large Display And Controls – With a back-light for improved reading. There are also large buttons for easy operation of the measuring instrument.

Sensors/Calibration – The Rotronic CP11 features the Hygromer IN-1 humidity sensor. Calibration is possible at 35/80 %RH and calibration of CO2 possible at 0 and 400 ppm.

Rotronic CP11 CO2 Measurement Meter

Principles Of Operation

Rotronic CP11 handheld instrument evaluates the quality of air along with a measurement of the levels of carbon dioxide (CO2) gas present. In closed areas with inferior ventilation and populated with people, carbon dioxide levels can quickly develop in high concentrations.

CO2 gas is colourless and odourless making it impossible for humans to detect alone. In addition the gas occurs naturally in the earth’s atmosphere and a high CO2 content becomes apparent in humans through rapid fatigue and loss of concentration.

In order to initiate suitable counter measures such as an increase in the supply of fresh air, it is important to data log and record not only parameters such as relative humidity and temperature measurement, but also the CO2 gas content.

Rotronic CP11 CO2 Gas Measurement


Rotronic CP11 CO2 measurement instrument is supplied with SW2.1 (settings, data download, display of measured values) or compatible with optionally available HW4 software.

 

Rotronic CP11 CO2 GAS Meter – General Information

Type Parameters  Range of application Power supply IP Protection Clock Alarm
Handheld device CO2 gas, relative humidity and temperature 0…50 °C / 0…100 %RH, non-condensing 4 x AA batteries / optional AC adapter AC1214 IP30 Real time clock Adjustable for CO2 measurement

Technical Information

Battery life Current consumption Warm-up time Calculations Measured value memory
50 h 40 mA (typical) <1 min. Dew point, wet bulb temperature 18,000 values with time stamp, automatic recording (%RH / °C / ppm) 99 single values with time stamp, manual recording (%RH / °C / DP / Tw / ppm)

CO2 GAS MEASUREMENT & Detection Information

Measurement principle Measurement range Accuracy at 23 °C ±5 K Response time Adjustment points Pressure dependence Null drift Maintenance
Non dispersive infrared (NDIR) with automatic baseline correction (ABC) 0…5000 ppm ±30 ppm ±5 % of the measured value <10 sec @ 30 cc/min. flow, <3 min diffusion time 0, 400 ppm +1.6 % reading per kPa <10 ppm/year No maintenance

 

Rotronic CP11 CO2 Measurement Instrument Meter Guidelines

350 – 450 ppm 400 – 1,200 ppm >1,000 ppm 5,000 ppm (0.5 %vol) 38,000 ppm (3.8 %vol) >100,000 ppm (10%vol)
Fresh air outdoors Room air Fatigue and loss of concentration become apparent. Maximum permissible value at the workplace during an 8-hour workday Breathing air (direct exhalation) Nausea, vomiting, loss of consciousness and death

 

ROTRONIC – CARBON DIOXIDE GAS MEASUREMENT FOR VENTILATION MONITORING

Rotronic devices including transmitters, data loggers and hand held instruments are used for regulating ventilation in living areas, offices, classrooms, cinemas, hospitals and also for measuring CO2 pollution in garages, tunnels and logistics facilities. Rotronic carbon dioxide gas measurement instruments monitor indoor air quality (IAQ) safeguarding the health, productivity and comfort of occupants.

Carbon Dioxide Measurement - Rotronic

Image : Wired

ROTRONIC MEASUREMENT

View the current Rotronic catalogue :

HUMIDITY, DEW POINT, FLOW, PROCESS PRESSURE, CARBON DIOXIDE & TEMPERATURE MEASUREMENT

T&D distribute Rotronic probes, instruments, transmitters, data loggers, indicators and sensors for the measurement of dew point, flow, air velocity, process pressure, carbon dioxide and temperature. Click an icon for further information…..

Rotronic - Dew Point Measurement

Low Dew Point

Rotronic - Flow Measurement

Flow & Air Velocity

Rotronic - Process Pressure Measurement

Process Pressure

Rotronic - CO2 Measurement

Carbon Dioxide

Rotronic - Temperature Measurement

Temperature

 

Data Sheets